Ultra-high dose rate X-ray radiator for studying flash radiotherapy

Oct 15, 2026, 2:10 PM
20m
Conference Room 2, 4th floor (SARI, Shanghai, China)

Conference Room 2, 4th floor

SARI, Shanghai, China

Bldg. 10, No. 100 Haike Rd., Pudong District, Shanghai, China
Talk Accelerator technologies for industrial & medical applications Parallel B

Speaker

Hao Zha (Tsinghua University, Beijing, China)

Description

FLASH radiotherapy is a rapidly growing topic, with most evidence at the cell/animal level. It may spare normal tissue while maintaining tumor control comparable to conventional RT, but clinical progress remains slow. A major bottleneck is the lack of suitable accelerators/radiators. Protons from cyclotrons offer ultra-high dose rate (UHDR) and penetration but not rapid energy modulation; carbon synchrotrons lack UHDR; electron LINACs provide UHDR but limited penetration; X-ray LINACs offer penetration and need no energy modulation but currently lack UHDR. For protons/heavy ions, magnet tuning/switching takes 50–100 ms, still insufficient and lowering average dose rate in some tissues. Electrons can reach kGy/s–MGy/s, but conventional LINACs are 5–20 MeV; VHEE ≥100 MeV requires large, expensive facilities, though compact X-band/S-band designs are emerging. Laser plasma wakefield accelerators are compact but low dose rate.

X-rays were selected for narrow penumbra (10–90%: 6 mm), large fields, conformality, easier dosimetry/shielding, and compact economic potential (6–10 MeV, <1 m). Challenges include low dose rate (0.1–0.2 Gy/s vs FLASH threshold ≥40 Gy/s) and multi-angle irradiation. UHDR X-ray animal experiments use short SSD and single angle; clinical use needs large SAD and multiple angles. Gantry rotation within FLASH time is nearly impossible, so static multiple irradiators are proposed: one accelerator with multiple beamlines or multiple compact accelerators. Reported approaches include LLNL’s induction accelerator with four parallel beams, UCLA/RadiaBeam fast-rotating VMAT-FLASH, and CAEP’s superconducting accelerator, but 15 Gy/s at SAD=50 cm remains insufficient.

Tsinghua developed a compact S-band BTW linac for UHDR X-ray. A standing-wave S-band linac achieved 10 MeV, 7.5 MW peak power, 431 mA measured pulse current, 1% duty cycle, and 72 Gy/s at SAD=70 cm. A clinical concept uses five linacs with carefully chosen angles, suitable for IMRT and rotatable across fractions. A 2025 prototype is 3 m diameter and 2.5 m long, fitting conventional bunkers, with 45 Gy/s at SAD=80 cm and planned upgrade to 100 Gy/s. Compactness and cost-effectiveness remain key advantages of X-ray FLASH-RT.